搜索

Product Gallery

Double Door Blood Bank Refrigerator 660L

CERTIFICATES:ISO, CE

WARRANTY:3years

MOQ:1unit

CHARACTERISTICS

ModelGBR-S110GBR-S210GBR-D660GBR-D1010
Size
Volume (L/cu.ft)110/3.88210/7.42660/23.311010\35.67
Capacity of Blood Bags(450ml)66132360576
N.T./G.T.(kg)50/5567/75165/195256/286
Exterior Size(W*D*H)(mm)522*600*1050522*600*15501220*648*18851220*878*1885
Interior Size(W*D*H)(mm)432*462*478432*462*9781100*454*13251100*684*1325
Package Size(W*D*H)(mm)572*685*1070572*685*15701300*730*20801300*960*2080
Performance
Temperature Range(℃)4士1℃
Temperature Accuracy(℃)0.1℃
Climate/ Ambient Temp.(℃)SN (10-32℃)
Noise(db)49.6dB(A)51.2dB(A)51.2dB(A)53.6dB(A)
Refrigeration System
RefrigerantCFC-free, R134a
CoolingForced-air cooling
DefrostAuto
Compressor Brand/QtyEMBRACO/1
InsulationPURF
Control System
Temperature ControllerMicroprocessor
SensorNTC
DisplayDigital LED display
Power
Voltage Frequency(V/Hz)220V, 50Hz/60Hz
Power Consumption(kWh/24h)1.952.226.318.53
Power(W)160200364513
Current(A)0.60.82.422.37
Construction
Standard Shelf/drawer
APS: Anti-corrosion Painted steel
SS: Stainless steel
2 APS Shelves
6 APS Baskets 
4 APS Shelves
12 APS Baskets 
2 APS Shelves
36 APS Baskets 
12 SS Drawers
Inner Door Qty2466
Inside MaterialStainless steel
Outside MaterialAnti-corrosion painted steel
Outer Door Glass1 Double-layer glass door with heating film2 Double-layer glass door with heating film
CasterStandard
Audible and Visual Alarms
High/Low TemperatureStandard
Temperature Controller FailureStandard
Sensor FailureStandard
Door AjarStandard
Power offStandard
Backup Battery MalfunctionStandard
Accessories
Backup Battery 72h for alarm if power off
Test Hole1/25MM
Thermal PrinterStandard
Drainage SystemStandard
LightAuto LED light
Safe AccessDual guard with door lock and password protection
Temperature RecorderOptional


DESCRIPTION

APPLICATIONProfessional for storage of blood,serum, as well as medicines and bio products.
Suitable for blood banks, hospitals, CDCs
ERGNOMIC DESIGN• Caster design and interior LED light
• Safety lock to prevent unauthorized access
• Storage space designed for easy sorting of a variety of blood products.
RELIABILITY & KEY FEATURE• Consistent cabinet temperature 2-6℃
• Auto defrost to remove moisture on cooling surface.
• Forced-air cooling with electrical heat compensation system.
• LED display with microprocessor controller of o.1℃ resolution.
• High-tech integrated sensors to test and control temperature.
SAFETY•  Build-in backup battery to display temperature and tooperate audible an visual alarm systems for up to 72 hours without ac power.
•  Equipped with a complete temperature alarm system featuring audible buzzer and visible flashing light.
•  Five alarm conditions: High/low temperature, power failure, voltage abnormal, door ajar, backup battery.
•  Standard with acrylic inner small door, for less affection on other room when one door open.
•  Standard with laser printer.

Double Door Blood Bank Refrigerator 660L(图1)

VIDEO

CARALOG

Related Products